PEER-REVIEWED RESEARCH. A label-free electrochemical sensor measured synthetic miR-181b across a six-order-of-magnitude range: 1 fM to 1 nM, with an experimental detection limit of 1 fM. Chen and colleagues combined a peptide-nucleic-acid capture layer, exponential amplification reaction (EXPAR), and a ruthenium/iron catalytic readout. At 1 nM equimolar tests, the assay distinguished miR-181b from closely related miR-181a, miR-181c, and miR-181d, although miR-181d still produced a smaller cross-signal. Nine electrodes tested across three days gave 4.34% relative standard deviation; electrodes retained 83.8% of their initial response after six days in PBS at 4 °C. The boundary matters: every quantitative test used synthetic target, not serum or clinical specimens, and long-term shelf life and inter-batch variation remain untested. Source: Biosensors. Find the paper by searching the exact title “Label-Free Peptide Nucleic Acid-Based Electrocatalytic Biosensor Coupled with EXPAR for Ultrasensitive Detection of miR-181b.”