PREPRINT (not peer-reviewed). A small automated line made and screened 173 powder blends—but cleaning proved a major limitation in one autonomous run. Tsioutsios and colleagues linked vacuum-assisted dosing, pneumatic blending, robotic sampling and near-infrared analysis. A relative spectral-dispersion screen flagged five of the 173 binary and ternary blends as atypical; this is not direct proof of absolute homogeneity. In paracetamol tests, predictions tracked targets from 10–30 wt.% with within-blend variation of 2.0–3.7% at 15–30 wt.%; at 35 wt.% the relative error was −18.37%. Twelve formulations took about six hours in an autonomous sequence, where increasing carryover exposed a cleaning limitation. Source: ChemRxiv. Find the paper by searching its exact title: An Automated Multi-Material Powder Dosing, Blending, and Formulation Evaluation Platform.