Post

PREPRINT (not peer-reviewed). In a nanogranular palladium hydrogen sensor, changing temperature can flip whether hydrogen raises or lowers resistance. Alkazaz, Zierold, and Huber tested an atomic-layer-deposited Pd network on macroporous silicon in 0.05–5 vol% H2/N2. Their two-channel model separates electron scattering inside particles from lower junction resistance as hydrogen-driven expansion narrows gaps. For the annealed device, the modeled zero-response point moves from 0.12 vol% H2 at 25 °C to 1.36 vol% at 40 °C; the 55 °C value is extrapolated beyond the measured range. At 55 °C, measured response stays negative but passes through a minimum near 0.3 vol%, so one late resistance reading can match two concentrations. A 10-minute, 200 °C air anneal also changes calibration and reverses the low-concentration room-temperature response. These are setup- and thermal-history-specific results, not a universal calibration rule. Source: arXiv. Find the paper by searching the complete title “Temperature Flips the Sign of the Hydrogen Response in Nanogranular Palladium Networks.”

6:43:59 AM · Nov 28, 2026