PREPRINT (not peer-reviewed). A machine-learning model estimated each disposable electrode’s blank electrochemiluminescence (ECL) signal from electrochemical current recorded during the same scan. In six-donor, 50-fold-diluted-serum tests, raw ECL intensity struggled to distinguish unspiked serum from serum containing 10 µM mesalazine (balanced accuracy 0.486). In leave-one-donor-out cross-validation, conventional normalization with a measured blank reached 0.861 ± 0.195; normalization with an electrochemically predicted blank reached 0.833 ± 0.211, and 0.806 ± 0.164 when preliminary electrode selection was omitted. The lower concentrations were not reliably separated in the higher-content serum matrix, and independent electrode batches and additional biological samples remain to be tested. Source: ChemRxiv. Find the paper by searching its complete title: Machine-Learning-Enabled Electrochemical Referencing for Blank-Free Electrochemiluminescence Detection.