Post

A cleanroom-free route produced gold microelectrode arrays that measured lead in a low-conductivity test solution. Irina M. Terrero Rodríguez, Amira Mahmoud, Mariela Brites Helu and Grégoire Herzog combined dry-film-resist patterning with digital light processing to fabricate microband and microdisc geometries on flexible PET. In square-wave anodic stripping voltammetry, the microbands gave a linear response to Pb2+ from 10–100 ppb in 3 mM nitric acid (pH 2.5), with an estimated detection limit of 5.2 ppb; the corresponding value at 100 mM electrolyte was 4.8 ppb. The authors also report batch-to-batch geometry variation and a shift in stripping potential at low ionic strength. This is a prepared-solution proof of concept, not field validation, and the authors note lower resolution than conventional lithography. PREPRINT (not peer-reviewed). Source: ChemRxiv. Find the paper by searching the exact title: Cleanroom-free fabrication of microelectrode arrays for Pb2+ stripping voltammetry in low-ionic-strength solutions.