PREPRINT (not peer-reviewed). Can a sensor tune its own measurement settings? In this ChemRxiv study, an automated Bayesian-optimization loop selected frequency, amplitude and step size for a pulsed electrochemical measurement on individual aptamer electrodes. For kanamycin sensors, it searched 50 settings rather than the 200-setting comparison sweep; on planar gold, the authors report mean quality-score improvements of about 46% ± 29% for signal-on and 37% ± 67% for signal-off responses. Automated titrations compared optimized settings with a fixed 200 Hz, 36 mV, 2 mV reference, with prediction accuracy varying by electrode morphology and response direction. In a separate nanoporous-gold vancomycin test, the reported RMS concentration-prediction fold error fell from 3.84× to 1.93×. These are setup-specific results—not proof of universal or clinical performance; broader biofluids, electrode sizes, instruments and sensor aging remain future work. Find the paper on ChemRxiv by searching its exact title: “Autonomous Bayesian Optimization in Electrochemical Aptamer Sensors.”